Quantitative and Qualitative Analysis of Total Oil and Grease in Water (ASTM D7575) using Agilent FTIR Cary 630 and ClearShot Extractor

The discharge of oil and grease into water environments such as oceans and rivers is a water quality control issue since it may lead to environmental problems including adverse effects on ecological systems and malodor.

Agilent FTIR
Request Info/Quote

What is ASTM D7575?

Standard method for determining total oil and grease in water, using FTIR and solvent free extraction by clearshot extractor.

ASTM D7575 Key Benefits

  • No Solvent required for the extraction, No Need for Chemical Hoods, Minimizes Glassware and Lab Space.
  • Rapid and simple samples extraction
  • Low sample analysis cost
  • Accurate and precise results
  • Ability to identify the specific oil or type of oils in the sample
  • Supplying certified Solid-state Calibration Standard Devices (CSDs) to prepare your calibration curve.
  • Cover wide range of TOG from 0.2 till 1000 ppm.

Cary630 Key Benefits

  • Mid-IR Spectrometers with special analysis to carryout heavy workload and covering different samples fields.
  • Easy to carry for using outside the lab. and field location
  • Interferometer: 25 mm, permanently aligned, Michelson, 45°, mechanical flexure
  • Enclosure: Sealed and desiccated optics system
  • Communication by USB-2
  • Size: 16 x 31 x 13 cm (6 x 12 x 5 inches) (W x D x H) and Weight: 3.8 kg (8 lbs.)
  • Power: 110 – 240 V AC, 60/50 Hz
  • Spectral range: KBr 7000–350 cm-1 and ZnSe 5100–600 cm-1
  • Sample interface and accessories: DialPath (30 to 250 µm), TumblIR, Diamond ATR, Diffuse Reflectance, Transmission, Germanium ATR, 5 Bounce ZnSe ATR, 10 Degree SRA, 45 Degree SRA

Sample Extraction Steps

  • Collect and acidify your samples at PH2, put the sample in Ultrasonic and heated to 40 °C for 20 minutes. then collect 10 ml sample by syringe.
  • Connect syringe to clearshot extractor, the extractor will trap oil and grease inside its membrane and release water.
  • Introduce the extractor with oil sample to FTIR card holder and measure by FTIR.
  • Open TOG quant method in Microlab software and measure your sample at 2920 Cm-1.

Results

  • Very accurate and reliable results in very short time
SampleAbsorbanceConcentration [ppm]
CSD10.00200.73
CSD20.0264.54
CSD30.09615.23
CSD40.2840.56
CSD50.6087.57
CSD60.87127.55
CSD71.2168.70